LEYBOLD-HERAEUS ULTRATEST M2/M2DS/M2BYリークテスター

ID:381307
ブランド:LEYBOLD-HERAEUS
モデル:ULTRATEST M2/M2DS/M2BY
タイプ:リークテスター
年:1985
場所:Colorado
アメリカ合衆国
在庫No.:392205
価格:

スペック

Equipped with (buyer to verify details:
Leybold-Heraeus NT 150/360 Frequency Converter (see below for details)

GENERAL SPECIFICATIONS FOR MACHINE AS FOLLOWS:
Smallest Detectable Leak
w/Options Quick-Test: 10-6 atm cc/sec
w/out Quick-Test: 2x10 -10 atm cc/sec
w/out Quick-Test w/reduced
Pumping Speed: 2x10 -11 atm cc/sec
Full Scale Leak-Rate Meter Reading
w/out Quick-Test: 10x10 -9 to 10x10 -6 atm cc/sec
w/Quick-Test: Must multiply meter reading by correction
factor (see manual Section 4-3-2)
Maximum Safe Operating Pressure
In the Mass Spectrometer (PIII) 1x10 -4 mbar
Detectable Masses: 3 and helium 4
Adjustable Emission Current: 0.6 to 5mA
Total Pressure Reading
Thermovac (PII): 10 -3 to 1000 mbar
Ionization Gauge (PIII): 10 -6 to 10 -3 mbar
Maximum Gas Flow
@ Test Port: 2x10 -3 atm cc/sec
Pumping Speet @ Test Port and In Mass Spectrometer
Ultratest M2: For Helium 20 ltr/sec
For Air 7.5 ltr/sec
Ultratest M2DS/M2BYTC/M2BYPC: For Helium 16 ltr/sec
For Air 6 ltr/sec
Backing Pump: Trivac Vane Pump
M2 and M2DS: D2A (2.2cfm or 1.1 ltr/sec displacement)
M2BYTC/M2BYPC: D4A (4.5cfm or 2.1 ltr/sec displacement)
Pump Vluis for Trivac: HE-200
Diffusion Pump: Leybodiff 180L
Pumping Speed For Air: 180 ltr/sec (381cfm) at 10 -5 mbar;
throtted to approx. 2 ltr/sec
Pump Fluid For Diff. Pump: HE-300 (50cc optimum)
Cold Trap
Capacity: 2.4 qts of liquid nitrogen (LN2)
Holding Time: Approx 20 hrs w/one Filling
Outputs - Recorder Socket But
Pin 5 (+) to Pin 3: 0 - 10 VDC (10 VDC is full scale
deflectio of leak-rate meter)
Pin 1 (+) to Pin 3: 1 - 7 VDC in 1 volt steps corresponding
to the 7 leak-rate positions on the
range switch. The output is 0 when the
range switch is not in a leak-rate
position.
Triggers 1 & 2 (Optional): 6 Amp, 220 VAC max.
Mains Supply: 115V, 60 Hz
Recommended Mains Fuse: 10 Amp
Dimensions
PK Unit: 22.8" x 16.3" x 16.5"
Electronic Control Cabinet: 9.2" x 10.3" x 16.1"
FB Remote Control Module: 8.7" x 9.1" x 5.1"
Optional Quick-Test Housing: 8.4" x 6" x 3.2"
Weights
PK Unit:
Ultratest M2: 154 lbs
Ultratest M2DS: 165 lbs
Ultratest M2BYTC/M2BYPC: 176 lbs
Electrical Control Cabinet: 28.7 lbs
FB Remote Control Module: 4.4 lbs
Optional Quick-Test Housing: 5.5 lbs
Data for Optional Quick-Test
Smallest Detectable Leakrate: 10 -6 atm cc/sec
Gas Throughput in Probe: Approx 90 atm cc/minute
Sampled Gas Inlet to Leak
Detector: Approx 10 - 3 atm cc/sec
Connecting Flange: NW40KF
Connecting Flange for
Calibrated Leak: NW10KF
Line Voltage: 60 Hz, 220 V
Dimensions of Housing: 8.4" x 6" x 3.2"
Weight of Housing: 5.5 lbs
Probe PVC Tubing
Length: 15' standard, up to 100' optional
Response time for 15' tubing: Less than 2 seconds
Weight: 1.1 lbs

GENERAL SPECIFICATIONS FOR FREQUENCY CONVERTER AS FOLLOWS:
Power Source
Voltage: Single phase, 100/120/220/240
Frequency: 47-63 Hz
(Currently set up for 120V / 60Hz)
Steady Rate Max: 680 VA
Transient (<0.5 Sec): 750 VA
Output Values
Voltage: 45 Vrms
Rated Frequency:
TMP-150: 840 Hz
TMP-360: 760 Hz
Max. Power:
Acceleration: 380 Watts
TMP-150 Normal Operation: 112 Watts
TMP-360 Normal Operation: 260 Watts
Relay Contacts:
Maximum: 24 VDC & 240 VAC @ 8 A
Mininum: 5 VDC @ 1 mA
Environment
Operating Temperature Range: 0 deg to 45 deg C
Storage Temperature Range: -40 deg to +85 deg C
Mechanical Design
Dimensions approx (w x h x d): 8.39" x 5.08" x 11.42"
Weight approx: 18.7 lbs

詳細

LEYBOLD-HERAEUS, Ultratest, No. M2, M2DS, M2BY, Helium Leak Detector

このモデルの詳細

の詳細について表示するには、 LEYBOLD-HERAEUS ULTRATEST M2/M2DS/M2BYレビュー、ブランド情報、写真、ビデオ、および多くを含むが、 ここをクリック