CINCINNATI SUB-ZERO ZP-32エンバイロメンタル チャンバー

タイプ:エンバイロメンタル チャンバー


チャンバー-幅:965.2 mm
チャンバー-長さ:965.2 mm
チャンバー-高さ:965.2 mm
最大 温度:190.6°C
大きさ:49.6"W x 70.8"D x 91.5"H

Cincinnati Sub-Zero Temperature Humidity Test Chamber – Model ZPH-32-1.5-H/AC

Temperature Range: -34°C (-30°F) to +190°C (+375°F)

Humidity Range: 10% to 98% relative humidity as limited by a +85°C (+185°F) maximum dry bulb and a +7°C (+45°F) minimum dew point.

Control Stability: ±0.5°C (+1°F); ±3% relative humidity at steady state conditions after stabilization

Temperature Performance: The ZPH-32-1.5-H/AC will cycle between +85°C (+185°F) and
(Non-humidity operation) -20°C (-4°F) at an average temperature change rate of 2°C per minute based on an empty chamber.

Temperature pull down from +24°C (+75°F) to:
-18°C ( 0°F) in 20 minutes
-34°C (-30°F) in 55 minutes

Temperature heat-up from:
+24°C (+75°F) to + 94°C (+200°F) in 20 minutes
+24°C (+75°F) to +190°C (+375°F) in 60 minutes
-34°C (-30°F) to + 24°C (+ 75°F) in 10 minutes

(Based on +24C/+75F ambient, 230 volt service,
and use of included accessories. Some accessories
may affect performance.)

Live Load Capacity: 1,200 watts at -18°C ( 0°C)
250 watts at -34°C (-30°F)
Note: Humidity control will be affected by the addition of live load.

Workspace Volume: 32 Cubic Feet (905 Liters)

Interior Dimensions: 38" wide x 38" deep x 38" high
(97cm wide x 97 cm deep x 97 cm high)

Exterior Dimensions: 46" wide x 83.75" deep x 89.50" high
(117 cm wide x 213 cm deep x 227 cm high)
(Addition of certain accessories may increase dimensions.)

Shipping Weight: Approximately 2,075 pounds

Utility Requirements: 208 or 230 volt (±5%), 60 hertz
- 1 phase, 28 FLA, 35 amp minimum service
- 3 phase, 26 FLA, 35 amp minimum service
(Customer to provide power cord and plug)

Heat Rejection: Approximately 6,400 watts (22,000 BTU/hr) maximum during cooling operation.

Air Flow